The COMET M1220 is the universal and network-capable multilogger with 2× MiniDIN inputs for temperature-humidity probes (Digi/M series) and 2× thermocouple inputs (type K, J, S, B, T, N) – the professional all-in-one solution for combined room climate and process temperature monitoring in climate chambers, warehouses, cold rooms, server rooms, incubators, dryers, ovens, burners, pharmaceutical production, food industry, or anywhere where temperature, humidity, and high temperatures need to be recorded simultaneously. With a huge memory of 1 million measured values (non-cyclic) or 600,000 values (cyclic), integrated Ethernet port for permanent network connection, backlit LCD, and traceable calibration certificate (EN ISO/IEC 17025), it is perfect for remote monitoring, quality management, audits, ISO compliance, and GMP applications.

The 2 MiniDIN inputs accept temperature-humidity probes Digi/M (measurement range & accuracy according to probe, typically -30 to +105 °C / 0–100 %RH, dew point automatically calculated ±1.5 °C). The 2 thermocouple inputs support type K (-200 to +1300 °C), J (-200 to +750 °C), S (0 to +1700 °C), B (100 to +1800 °C), T (-200 to +400 °C), N (-200 to +1300 °C) with accuracy ±0.3 % of reading +1.5 °C (or +1 °C for type B above +300 °C). The logger stores up to 1 million values with freely adjustable interval (1 s to 24 h). The backlit LCD displays all channels simultaneously. Thanks to the Ethernet port, measured values can be viewed live in a web browser, stored, and downloaded – remote access, real-time monitoring, and automatic data transfer are possible. Data evaluation via free COMET Vision software – with tables, graphs, PDF, and Excel export. Acoustic/optical alarm on limit exceedance. The multilogger is ready to use immediately – Plug & Play for probes and thermocouples.

The multilogger is maintenance-free, long-term stable, and equipped with a traceable calibration certificate (EN ISO/IEC 17025) – highest quality for GMP, ISO, and audits. Battery life up to several months (depending on configuration) or permanent network operation via Ethernet. Whether as a stationary logger in a control cabinet with network connection or as a portable device with battery – the M1220 offers maximum flexibility and reliability. The combination of 2× T+RH and 2× TC allows mixed operation – ideal for complex measurement tasks with room climate and process temperature.

Why the COMET M1220 is the best multilogger with 2 T+RH + 2 thermocouple and Ethernet:

  • 2× MiniDIN (T+RH) + 2× thermocouple – combined room climate & process temperature measurement
  • Wide measurement range – TC up to +1800 °C (type B), T+RH according to probe
  • 1 million measured values memory – non-cyclic or cyclic
  • Integrated Ethernet port – remote access, web browser view & network logging
  • Backlit LCD – all channels readable simultaneously
  • Acoustic/optical alarm – immediate warning on limit exceedance
  • Min/Max memory – extreme values per channel
  • Traceable calibration certificate included – EN ISO/IEC 17025
  • Autonomous battery operation or mains operation – maximum flexibility
  • Free COMET Vision software – graphs, PDF & Excel export
  • 3 years warranty – COMET quality guaranteed

Technical Data M1220:

  • External Pt1000 temperature: -200 to +600 °C (±0.2 °C from -200 to +100 °C / ±0.2 % above)
  • External T+RH (Digi/M): according to connected probe
  • Dew point: according to connected probe (±1.5 °C at T<25 °C and RH>30 %)
  • Thermocouple type K, N: -200 to +1300 °C (±0.3 % +1.5 °C)
  • Type J: -200 to +750 °C (±0.3 % +1.5 °C)
  • Type S: 0 to +1700 °C (±0.3 % +1.5 °C)
  • Type B: 100 to +1800 °C (±0.3 % +1 °C above +300 °C)
  • Type T: -200 to +400 °C (±0.3 % +1.5 °C)
  • Resolution: 0.1 °C
  • Channels: 2× MiniDIN (T+RH) + 2× thermocouple
  • Memory: 1,000,000 values (non-cyclic) / 600,000 values (cyclic)
  • Recording interval: 1 s to 24 h
  • Measurement interval / display update: 0.7 to 5 s
  • Alarm: acoustic + optical (adjustable limits)
  • Display: backlit LCD
  • Communication: Ethernet port + USB
  • Power supply: battery or power adapter
  • Operating temperature: -10 to +60 °C
  • Protection rating: IP20
  • Warranty: 3 years

Perfectly suited for:

  • Combined room climate & process temperature monitoring
  • Climate chambers & incubators
  • Warehouse and cold room monitoring
  • Pharma and food production
  • Server rooms and data centers
  • Oven & heat treatment systems
  • Remote monitoring via Ethernet
  • Quality management & audits

Installation and use – done in minutes

Connect probes/thermocouples, plug in Ethernet cable or activate battery – ready! Set intervals and limits, start logging – read live data in browser or via COMET Vision. The backlit display and acoustic/optical alarm make operation safe and comfortable. Calibration certificate included – ideal for audits and quality management.

Summary:

  • Universal data logger with 4 inputs
  • Two inputs for temperature and humidity probes and two inputs for thermocouples K, J, S, B, T, N
  • The multilogger can be permanently integrated into an Ethernet network or used as a mobile device
  • A traceable calibration certificate according to EN ISO/IEC 17025 is included in the scope of delivery
  • The multilogger is a battery-operated data logger that records sensor and electrical measured values via its inputs
  • The data logger can be connected to an Ethernet network
  • The measured values can then be displayed in a web browser, stored, and downloaded for later analysis.

Scope of delivery:

  • M1220
  • Instruction manual
  • Traceable calibration certificate
  • 3× rechargeable AA batteries
  • Power adapter
  • USB cable
  • Wall mount
  • Free full version of COMET Vision analysis software
  • External probes are not included in the scope of delivery
Datasheets
Datasheet 1

Oops! Your browser doesn't support PDFs!

Datasheet 1 (2.54 MB)
Datasheet 2

Oops! Your browser doesn't support PDFs!

Datasheet 2 (151.95 KB)

Multilogger – Thermo-Hygrometer with 2 MiniDIN and 2 Thermocouple Inputs

Measured Quantities

  • Temperature
  • Relative Humidity
  • Calculated Values (Dew Point, Absolute Humidity, Specific Humidity, Mixing Ratio, Specific Enthalpy)

Temperature Sensor – External Pt1000 Probe (MiniDIN)

  • Measurement Range: −200 °C to +600 °C
  • Accuracy:
    • ±0.2 °C (−200 to +100 °C)
    • ±0.2 % of Reading (+100 to +600 °C)
  • Resolution: 0.1 °C

Temperature & Humidity Sensor – External T+RH DIGI Probe (MiniDIN)

  • Measurement Range: Depending on the connected Digi/M probe
  • Accuracy: Depending on the connected Digi/M probe
  • Resolution: 0.1 °C

Dew Point

  • Measurement Range: Depending on the connected Digi/M probe
  • Accuracy: Depending on the connected Digi/M probe
  • Resolution: 0.1 °C

Temperature Sensor – Thermocouple Type K, N

  • Measurement Range: −200 °C to +1300 °C
  • Accuracy: ±0.3 % of Reading + 1.5 °C
  • Resolution: 0.1 °C

Temperature Sensor – Thermocouple Type J

  • Measurement Range: −200 °C to +750 °C
  • Accuracy: ±0.3 % of Reading + 1.5 °C
  • Resolution: 0.1 °C

Temperature Sensor – Thermocouple Type S

  • Measurement Range: 0 °C to +1700 °C
  • Accuracy: ±0.3 % of Reading + 1.5 °C
  • Resolution: 0.1 °C

Temperature Sensor – Thermocouple Type B

  • Measurement Range: +100 °C to +1800 °C
  • Accuracy: ±0.3 % of Reading + 1 °C (in range +300 °C to +1800 °C)
  • Resolution: 0.1 °C

Temperature Sensor – Thermocouple Type T

  • Measurement Range: −200 °C to +400 °C
  • Accuracy: ±0.3 % of Reading + 1.5 °C
  • Resolution: 0.1 °C

DC Voltage Measurement

  • Measurement Ranges: −60 to +140 mV; −18 to +18 mV
  • Accuracy: ±100 µV; ±20 µV
  • Resolution: 0.1 µV

General Technical Data

  • Operating Temperature: −10 °C to +60 °C
  • Channels: 4× external sensor inputs (2× MiniDIN, 2× Thermocouple)
  • Memory:
    • 1,000,000 readings (non-cyclic mode)
    • 600,000 readings (cyclic mode)
  • Recording Interval: 1 s to 24 h
  • Display and Alarm Update: Every 10 s
  • Recording Mode:
    • Non-cyclic – recording stops when memory is full
    • Cyclic – oldest data is overwritten
  • Temperature Units: °C or °F
  • Interfaces: Ethernet, USB, RS232
  • Ethernet Protocols: WWW, SNMPv1, XML, DATALINK
  • Alarm Protocol: E-Mail
  • Power Supply: AC adapter 230 VAC / 5 VDC (required for Ethernet operation)
  • Battery Life: Approx. 3–4 weeks
  • Protection Rating: IP20
  • Dimensions (without connectors): 178 × 95 × 37 mm
  • Weight (incl. batteries): Approx. 380 g
  • Warranty: 3 years
Skip product gallery

Compatible Temperature Probes

COMET CP500 – Type K Surface Probe (-65 to +500 °C), Subminiature Connector
COMET CP500 – Type K Surface Probe (-65 to +500 °C), Subminiature Connector
COMET CP500 Fast Response Type K Surface Probe -65 to +500 °C, Class 1, Subminiature Connector
Product number: CP500

Regular price: €72.00
COMET CZ550 – Type K Penetrating Needle Probe (-65 to +550 °C), Subminiature Connector
COMET CZ550 – Type K Penetrating Needle Probe (-65 to +550 °C), Subminiature Connector
COMET CZ550 Type K Penetrating Needle Probe -65 to +550 °C, t90 ≈ 3 s, Stainless Steel V4A, Class 1, Subminiature Connector.
Product number: CZ550

Regular price: €78.00
COMET CZ900 – Type K Penetrating Needle Probe (-65 to +1000 °C), Subminiature Connector
COMET CZ900 – Type K Penetrating Needle Probe (-65 to +1000 °C), Subminiature Connector
COMET CZ900 Type K Penetrating Needle Probe -65 to +1000 °C, Class 1, Subminiature Connector
Product number: CZ900

Regular price: €78.00
COMET GD1250-1 – Robust Type K Thermocouple (-65 to +980 °C)
COMET GD1250-1 – Robust Type K Thermocouple (-65 to +980 °C)
Select cable length: Cable length 1 meter
COMET GD1250-1 Type K thermocouple with Nextel insulation, -65 to +980 °C, Class 1, subminiature connector.
Product number: GD1250-1

Regular price: €37.00
COMET GD260-1 – Type K Thermocouple Probe (-65 to +260 °C), Subminiature Connector
COMET GD260-1 – Type K Thermocouple Probe (-65 to +260 °C), Subminiature Connector
Select cable length: Cable length 1 meter
COMET GD260-1 Type K Thermocouple, extremely fast, Teflon-insulated, Ø 0.8 mm, -65 to +260 °C, Class 1, Subminiature Connector.
Product number: GD260-1

Regular price: €21.00
COMET GD700-1 – Type K Thermocouple (-65 to +700 °C), Subminiature Connector
COMET GD700-1 – Type K Thermocouple (-65 to +700 °C), Subminiature Connector
Select cable length: Cable length 1 meter
COMET GD700-1 Type K thermocouple with glass braid insulation, -65 to +700 °C, Class 1, subminiature connector.
Product number: GD700-1

Regular price: €21.00
Skip product gallery

Compatible MiniDIN Temperature/Humidity Sensors

COMET DIGIH/M-1 – Temperature and Humidity Probe, -10 to +60 °C / 0–100 %RH, MiniDin Connector
COMET DIGIH/M-1 – Temperature and Humidity Probe, -10 to +60 °C / 0–100 %RH, MiniDin Connector
The COMET DIGIH/M-1 ultra-thin digital temperature and humidity probe, Ø 11 mm, with MiniDin connector, measures temperatures from -10 to +60 °C and 0–100 %RH.
Product number: DIGIH/M-1-1

Regular price: From €95.00
COMET DIGIL/M – Temperature and Humidity Probe -10 to +60 °C (±0.4 °C), MiniDin Connector
COMET DIGIL/M – Temperature and Humidity Probe -10 to +60 °C (±0.4 °C), MiniDin Connector
COMET DIGIL/M digital T+RH probe 54 x 18 mm with MiniDin connector measures -10 to +60 °C and 0–100 %RH.
Product number: DIGIL/M-1-1

Regular price: From €91.00
COMET DIGIS/M – Temperature and Humidity Probe, -10 to +60 °C, MiniDin Connector
COMET DIGIS/M – Temperature and Humidity Probe, -10 to +60 °C, MiniDin Connector
The COMET DIGIS/M digital temperature and humidity probe (63 x 16 mm) with MiniDin connector measures temperatures from -10 to +60 °C and humidity from 0 to 100% RH.
Product number: DIGIS/M-1-1

Regular price: From €83.00

Manufacturer Information:

COMET SYSTEM, s.r.o.

Bezručova 2901

756 61 Rožnov pod Radhoštěm
Czech Republic

Phone: +420 571 653 990

Email: info@cometsystem.com

Responsible person:

Didactum® Security GmbH

Im Mühlenfeld 4

48163 Münster
Deutschland

Fon: (+49) 250 19 71 63 54 / (+49) 171 33 11 577

Email: info@didactum-security.de

Questions?

You have questions about Comet products?

You would like a quote?

Contact us